Introduction |
The INFRATEK MODEL 31 HAND HELD, BATTERY OPERATED, SINGLE AND THREE PHASE POWER ANALYZERS are valuable engineering tools designed to offer precision measurement on all types of signals including electronic drives. |
Low cost - Easy to operate |
The Infratek Model 31 hand-held Power Analyzers are sophisticated measuring instruments, single phase and three phase, designed for laboratory and field use. Unlike other instruments in this price range the Model 31 Power Analyzers are suitable for measurements on frequency inverter drives, light ballasts, and other electronically controlled loads. The wide frequency range, the high common mode rejection, and the fact that you can measure currents of up to 50A directly without having to use clamps guarantees excellent accuracy for all types of measurements. |
The Infratek 31 Power Analyzers are controlled by six menu soft keys M1 through M6, five
cursor soft keys, and one power-on/off soft key. The operating procedure, to configure the display, to
configure the interface, to set scaling factors, and to select the input, the ranges, and many more functions
are self-explanatory. If you desire you can store your personal instrument configuration in nonvolatile memory
and have the unit start-up with your personal configuration at power-on. |
Model 31 High Speed Power Analyzers
| ||||||||||||||||||||
Accessories
| ||||||||||||||||||||
Options |
| Voltage | 7 ranges: 1V, 3V, 10V, 30V, 100V, 300V, 1000V | - | |
| Frequency range: DC-100kHz ¦ Coupling: AC/AC+DC | 1Hz-100kHz / DC-100kHz | ||
| Crest Factor 3:1 at 50% full scale ¦ Common Mode 50Hz / 100kHz | 130dB/70dB | ||
| Built-in star point network 500kΩ ¦ Maximum Input: Hi to Lo / Lo to case | 1000V / 600V | ||
|
Accuracy Grades: k=1, k=2, k=4 |
||
| Current | 10 ranges: 100mA,300mA,1A,3A,10A; 1,3,10,30,100A | Clamp: 1A - 1000A | ||
| Frequency range: DC-100kHz ¦ Coupling: AC/AC+DC | 1Hz-100kHz / DC-100kHz | |||
| Crest Factor 3:1 at 50% full scale ¦ Common Mode 50Hz / 100kHz | 150dB/100dB | |||
| 3A input: 3A cont./10A 5s; 50A input: 40A cont./50A 20s. | Ri=30mΩ / 3mΩ | |||
|
Accuracy Grades: k=1, k=2, k=4 |
|||
| Power | 70 Ranges corresponding to the products of voltage ranges time current ranges. | DC-100kHz |
| Accuracy 23°C ±3°C; 50A/Clamp input, 3A input 1Hz-1kHz Add accuracy percentage figures of current and voltage input 1kHz-100kHz Add accuracy percentage figures of current and voltage input, add ±2% (1-PF)/kHz of range |
PF = 0 to ±1 |
|
| Computed Values | Reactive Power, Var = ±(VA² - W²)½; Apparent Power: VA = Arms Vrms; Power Factor: PF = W/VA; Crest Factor: CF = Ap/Arms, Vp/Vrms; Maximum = Ap, Vp; Minimum: -Ap, -Vp; PtP: Maximum - Minimum; Impedance: Z = Vrms/Arms; Total Harm Dist: THD1 = (Irms² - IFund²)½/Irms THD2 = (I2_rms² + I3_rms² + In_rms²)½/Irms |
Add accuracy percentage figures of values involved in computation. |
| Mechanical Values | Total input power, Nm/s; Output power to load, Nm/s; Torque at axis of rotating machine, Nm; Speed, rpm; Efficiency: Eta = Output power / Input power |
Synchronous machines only. |
| Harmonic Analysis | Frequency range of fundamental ¦ 4Hz-50kHz | Harmonic 1-63 |
| Accuracy: harmonic current and voltage, same as rms current and rms voltage | - | |
| Computed values: harmonic power, harmonic phase angle (power factor) harmonic impedance. |
Add accuracy percentage figures of values involved in computation. |
|
| Frequency | 2Hz-100kHz; A; V: ±0.1% | - |
| Integrator | Energy, Accuracy: Wh, VAh; Basic accuracy of integrated quantity. | - |
| Data Logging | Output values, speed and duration is programmable; maximum speed: 10 values from a 3-phase system in 150ms. |
Range of signal frequency 5Hz to 1kHz |
| Current harmonics | Current harmonics in a 3-phase system are determined according to IEC1000-3-2 | - |
| Display Power/Battery Dielectric Strength Dimensions |
Display: Blue liquid crystal graphic display with FL backlight; - 64 x 120mm; - 128 x 240 pixels Charger: AC, 50-60Hz, 115/230V, 0.2AF, 10VA; sealed Lead gel battery, 6 hours Dielectric Strength: Current inputs to case; Voltage inputs to case; Line input to case Dimensions: H x W x D |
58x108mm; 128x240 pixels 3kV; 2kV; 3kV 50Hz/1 min 75x160x245mm; 1.5kg |
| Options | RS232 interface Windows operating software 95,98,NT,ME,2000,XP Current clamps 0-200A / 0-400A |
- |
Screen Shot - Measurement Mode (typ.) |
||||||||||||||||||||||||||||||||||||||||||
Screen Shot - Measurement Value Table |
||||||||||||||||||||||||||||||||||||||||||
Screen Shot - Setup Menu |
||||||||||||||||||||||||||||||||||||||||||