| INFRATEK | Switzerland | SM201 Spectral Multimeter |
|
Main Features
|
Display Features
|
DSP Processing Power
|
Interface - Input/Output
|
| SM201 Models | Specifications | Windows Software | Screen Shots | Datasheet |
Introduction |
The MODEL SM201 SPECTRAL MULTIMETER is a multipurpose instrument designed for laboratory and field use, and for production testing. The optional interfaces are supplied with opearting software which lets you read data, tailored to your needs, from SM201. High speed transfer to a computer and data storage in EXCEL are possible. |
Simple to use |
The SM201 is a high performance, high accuracy multimeter that allows you to measure current and voltage simultaneously and power for current wave forms whose fundamental is in the frequency range 20Hz to 1kHz. The SM201 exhibits a wide current range up to 30A. It can be extended up to 20000A to measure the rms current of a burst, for example. |
Extraordinary features |
For continuous signals, harmonics 1 to 63 are computed on-line. Pre-compliance testing for the main current harmonics is performed by selecting averaging time "IEC555-2". In the FLICKER opearting mode the SM201 will determine short time Flicker Pst and long time Flicker Plt. Through the use of the analog inputs and outputs the capabilities of the Spectral Multimeter are further enhanced. The SM201 reads eight signal inputs from tansducers and controls up to four analog outputs to drive chart recorders. |
Designed to last |
The SM201 is controlled by the menu soft keys M1 through M6 and the cursor soft keys. The operating procedure, to configure the display and the interfaces, to select the operating mode, to select the input, the ranges, and many more features is selfexplanatory. You can save your personal instrument settings and have the unit start-up with your personal configuration at power-on. For more informations about the Infratek Spectral Multimeter, please contact your nearest Infratek distributor |
SM201 Spectral Multimeters
| ||||||||||||||||||||
Accessories
| ||||||||||||||||||||
Options
|
| Voltage | Ranges, 8 ranges 2-6-20-sequence; 0.6V, 2V, 6V, 20V, 60V, 200V, 600V, 2000V | - |
| Frequency range | DC, 2Hz-300kHz | |
| Crest Factor | 4:1 at 50% full scale (fs) | |
| Input Impedance | >2.3MΩ | |
| Common Mode - 50Hz/100kHz | 155dB/95dB | |
| Standard accuracy 23°K ±3°K; rms, mean, rectified mean; for 0.6V range, input >50% fs 1Hz-1kHz = ±(0.1% rdg + 0.1% range) DC, 1kHz-10kHz = ±(0.2% rdg + 0.2% range), DC typical 10kHz-100kHz = ±(0.3% range +0.04%/kHz rdg) * 100kHz-300kHz = ±(0.3% range + 0.04%/kHz rdg), typical |
improved acc., 1Hz-400Hz ±(0.05% rdg + 0.07% range ) * 0.6V range typical |
|
| Current | Ranges, 10 ranges 1-3-10-sequence; 15mA,50mA,150mA,500mA,1.5A,5A; 1,3,10,30,100A | - | ||
| Frequency range | DC, 2Hz-300kHz | |||
| Crest Factor | 4:1 at 50% full scale (fs) | |||
| Common Mode - 50Hz/100kHz | 160dB/120dB | |||
|
improved accuracy 1Hz-400Hz ±(0.05% rdg + 0.07% range) DC typical ¹for 2 lowest ranges, input >50% fs |
|||
| Power | 80 Ranges corresponding to the products V x A | - |
| Frequency range of fundamental frequency | 20Hz-1kHz | |
| Accuracy 23°K ±3°K 20Hz-100Hz = 0.3% rdg + (0.3% + 1% sinΦ) range 100Hz-200Hz = 0.5% rdg + (0.5% + 2% sinΦ) range 200Hz-1kHz = 0.5% rdg + (0.5% + 0.2%/100Hz) range |
-80° < Φ < 80° -80° < Φ < 80° PF=1 |
|
| Frequency | 2Hz-100kHz, A or V triggered; Accuracy ±0.1% | 100% full scale |
| Computed Values | Accuracy; Reactive Power, Var = ±(VA² - W²)½; Apparent Power: VA = Arms Vrms; Power Factor: PF = W/VA; Crest Factor: CF = Ap/Arms, Vp/Vrms; Form Factor: FF = Arms/At, Vrms/Vt; Impedance: Z = Vrms/Arms; Total Harm Dist: THD = (Irms² - IFund²)½/Irms; Flicker Pst, Plt |
Add accuracy percentage figures of values involved in computation. |
| Integrator | Energy, Accuracy - Wh, VAh, Varh,; Basic accuracy of integrated quantity. | - |
| Harmonic Analysis | Frequency range of fundamental - 5Hz-60kHz | - |
| Range of harmonic | 1-63 | |
| Accuracy, Harmonic current and voltage 5Hz-1kHz = ±(0.2% rdg + 0.1% range) 1kHz-10kHz = ±(0.5% rdg + 0.5% range) 10kHz-60kHz = ±(0.7% range + 0.1%/kHz rdg), typical |
- | |
| Burst | Measures current - or voltage burst of an intermittent signal. | Minimum burst duration 1ms |
| Flicker | Determines short time flicker Pst and long time flicker Plt. | - |
| Display | Blue liquid crystal graphic display with FL backlight; - 64x120mm; - 128x240 pixels | - |
| Power | AC, 50-400Hz; Fuse: Power | 85V-240V; 2AF/20VA |
| Dielectric Strength | Inputs to case or power supply Line input to case |
2.5kV/50Hz/1 minute 1.5kV/50Hz/1 minute |
| Dimension | H x W x D; Weight | 150 x 235 x 320mm; 4kg |
| Options | IEEE-488-2, RS232, Centronics printer output 4 Analog outputs, Output impedance 100Ω accuracy 0.2% 4 Analog inputs, low range, input impedance 200kΩ accuracy 0.4% typical 4 Analog inputs, high range, input impedance 200kΩ accuracy 0.4% typical Rack mounting kit Humidity: KYG according to DIN 40040, max. 85% RH non-condensing |
0 - ±5V 0 - ±1V 0 - ±10V |